INTRODUCTION
Commercial synthetic polymers such as epoxy resin, polyacrylonitrile, polyacrylates, phenolic resin, polyethyleneglycol, polyurethanes, polyolefins, polysiloxanes and fluoropolymers, are widely used in many industry sectors. These polymers feature backbones with -C-C-, -C-O-, -C-N-, -C-S-, Si-O- bonds, with substituents such as alkyl, aryl groups, -OH, -NH2, etc., which may be further modified with F, Cl, and Br. Under heat treatment either in air, under inert or reduced pressure conditions, they may release variety of molecular species, including H2O, NH3, CO, CO2, HCN, HCNO, NOx, SO2, CSO, HF, HCl, HBr, CH4, C2H4, silanes, siloxanes, fluorocarbons, formaldehyde, phosgene, carbonyl fluoride and other volatile organics. These species not only lead to performance deterioration and ultimate failures, but also cause great safety and health concerns. As such, most of them are strictly regulated by EPA.
Mass spectrometric identification of these species generally requires vacuum conditions and identification quite often suffers from mass overlaps and poor ionization yields that makes detection and identification problematic. On the other hand, these molecular species have distinctive vibration patterns in the 450 cm-1 – 4000 cm-1 infrared regions (Figure 1). In fact, FTIR is the preferred method for analysis of combustion exhaust composition.1
TGA-FTIR combines the strength of TGA with FTIR for materials characterization and outgassing profiling. It accurately records the mass loss as a sample is heated up in the TGA analyzer, and identifies the released molecular species resulting from the corresponding mass loss, by flowing the released molecular species through a long optical path gass cell of the FTIR instrument. At Eurofins EAG Syracuse Lab the above TGA-FTIR setup assures high sensitivity detection up to 10-100 parts per million mass loss for 50-100 mg sample.
Figure 1 Reference FTIR spectra of common outgassing species
CASE STUDY
TGA-FTIR analysis is demonstrated with outgassing investigation of polyacrylonitrile (PAN). As carbon fiber precursor, PAN must go through a series of processes including thermo-oxidation stabilization, pyrolysis and graphitization, which are accompanied with intensive outgassing. TGA-FTIR provides a powerful insight to the pyrolysis kinetics and the outgassing profiles. Figure 2 depicts the time (temperature)-based outgassing profiles acquired on our TGA-FTIR instrument, for a thermo-oxidation stabilized PAN sample (Sigma-Aldrich), including CO, NH3, HCN, and HNCO, all highly hazardous air pollutants. The insert is the corresponding TGA profile, which shows the thermal decomposition pattern and pyrolysis kinetics. Figure 3 demonstrates the clear identification of the outgassing CO, NH3, HCN, and HNCO.
Figure 2 TGA-FTIR profile of thermo-oxidation stabilized PAN, showing outgassing of HCN, HNCO, NH3 and CO from 350 °C to 800 °C. Argon atmosphere. Heating rate 10 °C/min. Insert is the TGA profile.
Figure 3 Illustrative FTIR identification of outgassing species. Black – snapshot FTIR spectrum of outgassing at 500 °C; red – the reference FTIR spectra of HCN (a), NH3 (b), CO (c) and HNCO (d), respectively.
COMMON APPLICATIONS
STRENGTHS
LIMITATIONS
TECHNIQUE COMPARISONS
Residual Gas Analysis (RGA). RGA is a high vacuum outgassing technique operating under 10-7 – 10-9 torr. This technique is suitable to quantitatively assess a whole range of small molecules (up to 100 amu), including H2, moisture, O2 and N2, etc. RGA is commonly used to assess internal water vapor content, which is known to affect the reliability of packaged electronic, medical, optical and devices. Testing methods such as MIL-STD-883, MIL-STD-750, Method 1018 Internal Water Vapor Content must be strictly followed to meet the DLA lab suitablity requirement.
Thermal Desorption Gas Chromotography Mass Spectrometry (GCMS) is used to analyze organic contaminants outgassed from solid sample matrices. GCMS separates and identifies individual organic species, up to 1000 amu. EAG’s thermal desorption chamber (1 1/2” diameter x 4” long) can be heated at any temperature between 45°C and 300°C, normally for a time period of one to three hours. Volatile organic components will be detected and semi-quantitative results will be provided. Detection limits are as low as 10 ng/component. Thermal Desorption GCMS cannot easily detect low level moisture, atmospheric species or other small molecules below 40 amu.
TGA-FTIR AT EAG
EAG’s TGA-FTIR consists of a TGA instrument featuring a highly reliable horizontal dual-balance mechanism that minimizes buoyancy contributions to the underlying signal. Outgassed species are effectively transferred by carrier gas to a long optical pathlength FTIR cell, which is mounted on a Perkin Elmer® FTIR spectrometer. Both the transfer line and FTIR cell can be heated through separate temperature controllers.
SUMMARY
TGA-FTIR is widely used for composition analysis and failure analysis. Equipped with a heated, long optical path FTIR cell, EAG’s TGA-FTIR technique is sensitive for survey analysis of outgassing profiles, especially air pollutants, in the targeted temperature and atmosphere regime. We have helped many clients to assess the issues and take mitigating measures. In summary, EAG’s TGA-FTIR service features: